Page -- 18

HI-3604 Manual

one second followed by the stored value. Immediately upon

releasing the key, a new value may be logged by again pressing

the LOG pad. Up to 112 values can be stored in this memory.

When the memory is filled, successive operations of the LOG

key will store the most current reading in memory location 112.

PREV - Pressing the PREV key displays the last stored data

value. When the key is pressed, the identification number of the

value is first displayed for about one second followed by the

stored value. The stored value is displayed as long as the key is

depressed. About two seconds after releasing the key, the

display returns to the current reading. Successive operations of

the key will move the displayed value down towards the

beginning of the memory (value identification #1). If the PREV

key is operated while viewing value #1, the identification value

will "wrap around" to the highest stored identification number.

NEXT - Pressing the NEXT key displays the next value in the

data logging memory. If the NEXT key is operated while viewing

the last stored value, the identification number will "wrap

around" to value number one.

When the displayed value is below 5% of the current full scale,

the arrow at the left end of the BarGraph display will indicate.

When the "Below Range" arrow is indicating, the accuracy of

the reading may not be within the specified tolerance. When

possible, the scale should be changed to permit a normal field

strength reading without the "Below Range" indication. If the

displayed value is too high the " Over Range", the arrow at the

right end of the BarGraph, will indicate. Select the next

appropriate scale.