beautypg.com

ETS-Lindgren HI-3604 ELF Survey Meter User Manual

Page 24

background image

Page -- 16

HI-3604 Manual

To return to the AUTO RANGE mode, press and hold the SCALE

keypad until the AUTO indication is shown in the upper left area

of the LCD.

MAX - While using the instrument for field measurements, the

processor is continually saving the highest indicated reading. To

recall and display the highest reading, press the MAX keypad.

This maximum reading is indicated as long as the MAX keypad

is activated. The maximum reading is indicated by the MAX

indication near the right edge of the LCD. On releasing the MAX

keypad, the reading is held for about two seconds and then the

memory is cleared and a new maximum reading accumulated.

When the instrument is shifted between the electric and the

magnetic field mode, the MAX reading memory is cleared.

The MAX keypad is also used to adjust the H-field units of

measure. This is done just after the meter is turned on. During

the self-test routine, the display will show all segments lit.

When all the segments are lit, press and hold the MODE SELECT

keypad to enter the display setup mode. Hold the MODE

SELECT keypad until the display reads "F-#" and the current H-

field unit of measure. Press the MAX keypad to scroll through

the available units. Once the desired units are selected, press

the MODE SELECT keypad to save the current setup, leave the

setup mode, and begin normal measurement operation.

E/H - Pressing the E/H keypad toggles the operation mode

between Electric and Magnetic field indications. The current

measurement units are displayed to the right of the reading on

the LCD.

The E/H keypad is also used to adjust the display response time.

This is done just after the meter is turned on. During the self-

test routine, the display will show all segments lit. When all the

segments are lit, press and hold the MODE SELECT keypad to

enter the setup mode. Hold the MODE SELECT keypad until the

display reads "F-#" and the current H-field unit of measure.

Press the E/H keypad to scroll through the four display filter

settings. F-4 has the "fastest" response time, ie. the least