ETS-Lindgren HI-3604 ELF Survey Meter User Manual

Page 19

background image

HI-3604 Manual

Page -- 11

3.0

OPERATION

NOTE: The HI-3604 is enclosed in a rugged aluminum extruded

case for protection of its internal circuitry. Because of the

nature of its intended use, the field sensor extends from the

readout module. The structure of the sensor is a multilayered

printed circuit board. It is well secured internally to the

aluminum case and has a tough polyester covering but is

exposed to greater potential for physical damage because of its

size and location. Use care in handling the HI-3604 to avoid

damaging the sensor by striking it against objects or applying

excessive force to the sensor paddle. When not in use, keep the

HI-3604 in its protective case where the sensor paddle is

properly supported.

Start Up

Membrane switches are used for controlling the HI-3604's

operation. The switch keypads are activated by gently pressing

on the center of the pad with a finger tip. Do not use hard or

pointed objects to operate the switch keypads.

Turn the unit ON, the default measurement condition is

Magnetic Fields. The custom LCD readout displays the units of

measure in addition to the observed value. A Bargraph display

is provided along the top of the LCD window. This display is an

analog approximation of the currently displayed digital value as

a percentage of the full scale range. The Bargraph input is

unfiltered for fast response when searching for peak fields.

The digital display response is digitally filtered for increased ease

of operation. This smooths the response to rapidly changing

fields. In some measurement situations, however, it may be

helpful to increase the response of the digital display, ie. reduce

the response time. This is done when the instrument is turned

on. Refer to KEYPAD MATRIX in this section, and the operation

of the E/H keypad for details.