beautypg.com

Check feature, Making a measurement – ETS-Lindgren HI-1801 Microwave Oven Survey Meter User Manual

Page 19

background image

Operation

|

19

C

HECK

F

EATURE

The control switch has two positions, an “ON/10 MW” position and a
“CHECK” position. The “ON/10 MW” position is used for testing

microwave ovens in the normal manner.

In the “CHECK” position, the

meter will respond to very low levels of leakage and will assure you

that the instrument is responding to microwave energy allowing you to

check the operation of the instrument at any time including while

testing the microwave oven.

Meter readings in the “CHECK” mode have no significance; the only

purpose of this mode is to verify proper meter function..

1.

Hold the ON/OFF switch in the CHECK position and use the large

knob to adjust the meter to an indication between 0 and 1. Note that

the probe must be in zero microwave field with no RF energy present.

2.

Holding the ON/OFF switch in the CHECK position and the oven

operating, move the probe over the door and door gap surfaces of the

oven. In the CHECK mode, the sensitivity of the HI-1801 is increased

from five to ten fold.

Making a Measurement

Turn the oven on and proceed with the leakage measurement.

1.

Hold the probe by the red handle in one hand and hold the meter in the

other hand, with the ON/OFF switch in the

“ON/10 MW” position.

2.

Place the tip of the spacer cover against the oven surface, with the

probe handle perpendicular to the surface. While testing, hold the

meter away from the oven to minimize the possible RF pickup directly

from the oven to the meter case.

3.

Move the probe slowly, about one inch per second, keeping the cover

tip in contact with the oven and the probe handle straight.

4.

Once you have checked all around the door edges, around and across

the window, and at any louvers or vents, return to the spot where you

noticed the highest reading before.