beautypg.com

0 operation, Accuracy – ETS-Lindgren HI-1801 Microwave Oven Survey Meter User Manual

Page 15

background image

Operation

|

15

4.0 Operation

Before operating any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Do not attempt to operate a microwave oven

with:

Any object caught in the door

A door that does not close properly

A damaged door hinge, latch or

sealing surface

Do not use the HI-1801 without the spacer cone

in position.

The areas where leakage is likely to occur are around the door seal, the window,

and at ventilation louvers or vents. A damaged or improperly installed magnetron

mounting gasket may cause leakage in the control panel area of the oven.

Accuracy

The accuracy of the ETS-Lindgren Model HI-1801 Microwave Survey Meter is

within ± 1 db (+25%, -20%) when used according to these instructions:

1.

Always use a clean spacer cover. Accuracy will be affected by wear of the

cover and by dirt and metallic particles which may become imbedded in the

EPS (expanded polystyrene) foam. If the cover is damaged, contact

Customer Service for a replacement.

2.

Hold the instrument case at approximately 45º. The balance of the meter

movement causes the needle to move slightly when held horizontally (flat)

or vertically (upright). The meter is calibrated while at an angle of 45º and if

it is used at that angle, no additional error will be introduced.