20 |

Operation

The beaker must be placed in the center of the oven rack.

Close and latch oven door.

Set range timer to five minutes and select “High” (915 MHz) power.

When power indicator (or “Cook”) lamp lights, range is ready for test.

Testing the Range

Testing should begin immediately after the power indicator lamp

comes on. As you scan the perimeter of the oven door, watch the

meter for deflection. The entire perimeter of the oven must be

scanned.

Place the white probe cover adjacent to slot between door and oven

body, scan entire perimeter of oven door. Keep axis of probe

perpendicular to oven surface being measured.

Select “Fast” response for initial scan of entire door periphery to locate

the maximum leakage point. Meter readings noted on first scan should

indicate the suspect locations of radiation, if any exist.

Once the origin of the maximum leakage has been identified, turn the

select switch to “Slow” response and monitor for two minutes,

recording the maximum repeatable values obtained during this survey.