Pre-Installation Tasks

Probe Test

Turn the selection knob on the meter to the “PROBE TEST” position.

With the “ZERO ADJ.” knob in a center position, the meter should read

between the green lines (in the “OK PROBE TEST” area). If it does

not, the probe has been damaged and should not be used. A defective

probe cannot be replaced in the field and should be returned to

ETS-Lindgren for repair. See the Product Information Bulletin included

with your shipment for details regarding returning instruments for

repair.