Operation

| 19

The selection knob should be turned off when the meter is not in use.

Standard “Load” Versus Time

Laboratory tests indicate maximum radiation usually occurs within

three minutes after “Power On” with specified load. In order to measure

maximum radiation, scan perimeter of door when power comes on.

This initial scan will take approximately one minute and will locate the

maximum leakage point(s). Monitor the maximum leakage point(s) for

another two minutes.

Range Preparation

Move oven to edge of cabinet so that lower slot (door to cabinet)

overhangs.

Place 275 cc of cold water in 600 cc beaker and place beaker in center

of oven. Additional suitable beakers are available; see the

Replacement and Optional Parts table in the Maintenance Section of

this manual for more information.

Load specifications: 275 cc of cold tap water in a 600 cc beaker

(Kimax #14005 or equivalent). Do not deviate from the specified.