7.0

Setting the AUTO-OFF Function

You can set the amount of time the scale is inactive before it shuts down
to save battery power. Follow these steps:

1.

Turn the scale on. Press and hold down the Tare/Units key and the
On/Off/Zero key for more than 5 seconds until…

A.oFF.x” is displayed. X is the variable time in minutes. It can be
set from 0 to 9 minutes. If X=0, the AUTO-OFF function is
disabled.

2.

To set the value of X, press the Tare/Units key to scroll through the
choices.

3.

Press the On/Off/Zero key to accept the setting and restart the
scale.

8.0

Screen Messages

0¯¯¯¯¯¯:

Power-on zero-point is more than 20% of full capacity above
the calibration zero-point when scale is turned on.

OR

The weight reading is more than 5% of full capacity above the
power-on zero-point when you press the On/Off/Zero key;

0______:

Power-on zero-point is more than 20% of full capacity below
the calibration zero-point when scale is turned on.

OR

The weight reading is more than 5% of full capacity below the
power-on zero-point when you press the On/Off/Zero key;

_______:

the current weight is less than –25% of full capacity;

¯¯¯¯¯¯¯:

the current weight is over 102% of full capacity;

CAL-0:

Zero point at calibration;

CAL-F:

Full capacity at calibration;

CAL-Er: Calibration

error;

A.ch.E r: Analog

channel

error;

EEP. Er:

EEPROM accessing error or data in EEPROM is in error or
not in their normal range;

Lo.bAt:

The voltage is below 5.7V (±0.1V)

x.x Udc:

The voltage is x.xV (±0.1V)