0 setting the auto-off function, 0 screen messages – Salter Brecknell 405 User Manual
Page 10

10
7.0
Setting the AUTO-OFF Function
You can set the amount of time the scale is inactive before it shuts down
to save battery power. Follow these steps:
1.
Turn the scale on. Press and hold down the Tare/Units key and the
On/Off/Zero key for more than 5 seconds until…
A.oFF.x” is displayed. X is the variable time in minutes. It can be
set from 0 to 9 minutes. If X=0, the AUTO-OFF function is
disabled.
2.
To set the value of X, press the Tare/Units key to scroll through the
choices.
3.
Press the On/Off/Zero key to accept the setting and restart the
scale.
8.0
Screen Messages
0¯¯¯¯¯¯:
Power-on zero-point is more than 20% of full capacity above
the calibration zero-point when scale is turned on.
OR
The weight reading is more than 5% of full capacity above the
power-on zero-point when you press the On/Off/Zero key;
0______:
Power-on zero-point is more than 20% of full capacity below
the calibration zero-point when scale is turned on.
OR
The weight reading is more than 5% of full capacity below the
power-on zero-point when you press the On/Off/Zero key;
_______:
the current weight is less than –25% of full capacity;
¯¯¯¯¯¯¯:
the current weight is over 102% of full capacity;
CAL-0:
Zero point at calibration;
CAL-F:
Full capacity at calibration;
CAL-Er: Calibration
error;
A.ch.E r: Analog
channel
error;
EEP. Er:
EEPROM accessing error or data in EEPROM is in error or
not in their normal range;
Lo.bAt:
The voltage is below 5.7V (±0.1V)
x.x Udc:
The voltage is x.xV (±0.1V)