Troubleshooting

Table 1: Troubleshooting Guide

Symptom

Reason

Solution

No display when turned on

Battery dead or loose
contact

Check the battery
Tighten the contacts

Display flashes

Low battery

Recharge the battery

Scale won’t turn on or off

ON/OFF button damaged

Return unit for service

Scale won’t tare

TARE button damaged

Return unit for service

Recharge light does not
come on

The recharger may be
damaged or it is not
plugged into proper outlet

Check the recharger and
the connection to the AC
power source

Displayed reading changes

The load is unsteady

The loadcell is damaged

Humidity too high

Stabilize the load

Check and replace loadcell

Place unit in dry
envirornment

Scale will not show 0

Scale has not had sufficient
time to warm up after power
up

Scale has been off the hook
for too long

Let the scale stand for 3-5
minutes after power up
before weighing

Hang the scale and wait 3-5
mintues

Large weighing error

Scale is not hung properly.

Check the scale and the
shackle

Battery won’t recharge

Battery is damaged or outlet
is not working properly

Replace the battery or
check the outlet

Remote distance reduced

Receiver window obscured
by soil

Batteries low in remote.

Clean receiver window

Replace batteries