11

Access to the calibration switch may be sealed with a lead-wire combination

on the bottom of the scale. Prior to removing the lead seal, check to see if the

test switch is in the correct position by following step 1 below. If you are not

allowed access, remove the lead-wire seal.

Note: You can exit the calibration procedure at any time by pressing the On/

Off key.

1. To enter calibration mode, while in normal weighing mode, hold the Zero

and

On/Off keys at the same time until…

The scale beeps and, if the calibration switch is in the ON position,

the WEIGHT window shows CAL. oN. If the display shows CAL.oFF,

change the calibration switch position. See Figure

3

.

The PIECE WEIGHT window shows Unit. 1 or Unit. 0. If Unit. 1 is

displayed, the calibration unit of measure is pounds. If Unit. 0 is

displayed, the calibration unit of measure is kilograms.

2. Press the Units key to toggle between pounds and kilograms. When the

unit you want is displayed, press the

Enter key to accept it…

The scale beeps and unLoAd is displayed in the PIECE WEIGHT

window. A number representing raw A to D counts is displayed in the

WEIGHT window.

3. Remove all weight from the scale, wait for the scale to be stable (the

Unit

annunciator stops flashing) and press the

Enter key to capture the

zero point…

If the scale successfully captures the zero point, the scale beeps and

LoAd1 (load weight #1) is displayed in the PIECE WEIGHT window, it

means the scale will be calibrated on the 1st point weight.


4. Load first test weight that is > 10% of full capacity onto

the center of the weight platter, wait for the scale to be stable (the Unit

annunciator stops flashing) and press the

Enter key to capture the

span1 weight.

When the scale captures the span1, InP. Ld is displayed in the PIECE

WEIGHT window and 0.000 is displayed in the COUNT window. InP.

Ld stands for input load weight, input number of the loaded weight by

numerical keys. Press ENTER to confirm the input weight. The input

weight data will be displayed in WEIGHT window.

5.When LoAd2 (load weight #2) is shown in PIECE WEIGHT window, it
means the scale will be calibrated on the 2nd point, weight must
be weight #1 +10% or FS. If don't want to calibrate the scale on this point,
just press ENTER to go to step (6), otherwise; Load weight #2 onto the
center of weight platter and press ENTER to confirm the loaded weight
when the scale is stable and unit announciator is not flashing.