Installation, Operation remote alert replacement parts, Cont.) – Taylor-Wharton Low Level Alarm User Manual
Page 2
![background image](/manuals/628251/2/background.png)
Taylor-Wharton
4075 Hamilton Blvd.
Theodore, AL 36590
Phone: (334) 443-8680
(800) TW TANKS (898-2657)
Fax: (334) 443-2209
Duration of battery-powered operation can vary from
six months to one year depending on the usage. An
audible and visual alert will be activated once battery
power is low.
Installation
(cont.)
The Cryo-Storage low level alert automatically checks
the liquid level every 30 minutes. This conserves
battery power thereby extended the security. The 25
percent level notification provides sufficient time for the
user to refill the dewar.
The user can check the level without opening the
vessel by pressing the “LEVEL CHECK” button. Re-
sults will be displayed in 30 seconds. The possible
results are provided in the following table:
MUTE: The user can mute the audible
alarm at any time by pressing the “TEST/
MUTE” button. The alarm will remain muted
until the next 30 minute check is made.
However, the LED will remain red until the
LN
2
level is raised above 25 percent.
TEST: The status light and the audible horn
can be tested by depressing the “TEST/
MUTE” button. This test can only be made
so long as an alert condition does not exist.
The remote alert circuit is activated immediatley once
a Low Battery, Low LN
2
or Sensor Fault is detected.
The remote alert will remain active even if the battery
power is depleted. Pins 1 and 2 are closed in normal
operation. Pins 2 and 3 are closed during a remote
alert condition.
Operation
Remote Alert
Replacement Parts
Description
Part Number
Alarm Module…………………………………..5140-1124
Power Supply ………………………………….5190-3094
Remote Alarm Plug ………………………….R06K-8C20
Sensorcore Assembly for 18XT,20XT, or 34XT
.......................................................................R033-8C40
Sensorcore Assembly for 34HC …………....R034-8C40
Sensorcore Assembly for 35HC …………....R037-8C40
Sensorcore Assembly for 35VHC ………….R036-8C40
Sensorcore Assembly for 1575RS ………...RS15-8C40
Sensorcore Assembly for 3000RS ………...RS30-9C16
Sensorcore Assembly for 4800RS …….......R05K-8C27
Sensorcore Assembly for 3K …………..…...R03K-8C49
Ordering Information – Order all replacement parts
and accessories from your supplier or from Taylor-
Wharton at the addresses listed below. Please include
the part number, quantity, and description of each part
requested.
Conditions
Color
Flash
Rate
Audible
Level Ok
Green
30 sec.
No
Checking
Yellow
1 sec.
No
Low Battery
Red
1 sec
Yes
Low LN(2)
Red
2 sec.
Yes
Sensor Fault
Red
3 sec.
Yes
Remote Alarm Plug Connection