Interface and output, Maintenance and cleaning – Scilogex OS20-Pro User Manual
Page 11

9
Digital (Enhanced) Overhead Stirrer
•
Do not remove the power line when cleaning. Only use
recommended cleansers:
Dyes
Isopropyl alcohol
Construction materials Water containing tenside / Isopropyl
alcohol
Cosmetics
Water containing tenside / Isopropyl
alcohol
Foodstuffs
Water containing tenside
Fuels
Water containing tenside
Before using other method for cleaning or decontamination,
the user must ascertain with the manufacturer that this
method will not damage the instrument. Wear the proper
protective gloves during cleaning of the instrument.
Note:
•
Electronic device can not clean with cleanser.
•
If you require maintenance service, must be
cleaned the instrument in advance to avoid
pollution of hazardous substances, and to
send back into original packing.
•
If the instrument will not use for a long
time,please switch off and place in a dry,
clean, room temperature and stable location.
9.Interface and output
Overhead stirrer makes use of special accessories and cables
to connect external devices. Standard 9-pin interface is used
to be connected to the PC.
•
selected from the specified communication line of the
EIA-standard RS232C, corresponding with DIN66020.
•
Transmission method: Asynchronous signal transmission
in start-stop-operation.
•
Mode of transmission: Fully Duplex. 1 starts bit; 7
character bits; 1 stop bit.
•
Transmission speed: 9600 bit/s.
•
Data communication from laboratory instrument to
computer is only possible on demand of the computer.
Note:
D o n o t i n s e r t o r r e m o v e t h e R S 2 3 2 C
communication line when switch on!
10.Maintenance and Cleaning
•
Proper maintenance can keep instrument working
properly and lengthen its lifetime.
•
Do not spray cleanser into the instrument when cleaning.