7

Analog 7” Square (Hotplate) Magnetic Stirrer

6.Faults

Instruments can not be power ON

- Check whether the power cable is plugged in.

- Check whether the fuse is broken or loose.

Fault in power on self test

- Switch OFF the unit, then switch ON and reset the

instruments to factory default setting.

Temperature can switched on not reach set point

- Check whether the safety temperature value is set too

low.

Stir speed can not reach set point

- Check whether the stirring bar breakaway monitoring

is ON.

- Excessive medium viscosity may cause abnormal

speed reduction of the motor.

Heating or stirring function do not start when knob is
pressed

- Check whether the unit is in the mode setting

program.

- Exit from the mode setting and restart the heating/

stirring functions.

If these faults are not resolved,please set the instruments

to factory default setting or take the unit to technical service

center.

7.Maintenance and Cleaning

Proper maintenance can keep instruments working
properly and lengthen its lifetime.

Do not spray cleanser into the instrument when cleaning.

Unplug the power line when cleaning.

Only use recommended cleansers:

Dyes

Isopropyl alcohol

Construction materials Water containing tenside / Isopropyl

alcohol

Cosmetics

Water containing tenside / Isopropyl

alcohol

Foodstuffs

Water containing tenside

Fuels

Water containing tenside

Wear the proper protective gloves during cleaning of the
instrument.

B e f o r e u s i n g o t h e r m e t h o d f o r c l e a n i n g o r
decontamination, the user must ascertain with the

manufacturer that this method will not harm the

instrument.