Faults, Maintenance and cleaning – Scilogex MS7-H550-S User Manual
Page 9

7
Analog 7” Square (Hotplate) Magnetic Stirrer
6.Faults
•
Instruments can not be power ON
- Check whether the power cable is plugged in.
- Check whether the fuse is broken or loose.
•
Fault in power on self test
- Switch OFF the unit, then switch ON and reset the
instruments to factory default setting.
•
Temperature can switched on not reach set point
- Check whether the safety temperature value is set too
low.
•
Stir speed can not reach set point
- Check whether the stirring bar breakaway monitoring
is ON.
- Excessive medium viscosity may cause abnormal
speed reduction of the motor.
•
Heating or stirring function do not start when knob is
pressed
- Check whether the unit is in the mode setting
program.
- Exit from the mode setting and restart the heating/
stirring functions.
If these faults are not resolved,please set the instruments
to factory default setting or take the unit to technical service
center.
7.Maintenance and Cleaning
•
Proper maintenance can keep instruments working
properly and lengthen its lifetime.
•
Do not spray cleanser into the instrument when cleaning.
•
Unplug the power line when cleaning.
•
Only use recommended cleansers:
Dyes
Isopropyl alcohol
Construction materials Water containing tenside / Isopropyl
alcohol
Cosmetics
Water containing tenside / Isopropyl
alcohol
Foodstuffs
Water containing tenside
Fuels
Water containing tenside
•
Wear the proper protective gloves during cleaning of the
instrument.
•
B e f o r e u s i n g o t h e r m e t h o d f o r c l e a n i n g o r
decontamination, the user must ascertain with the
manufacturer that this method will not harm the
instrument.