9 associated standards and regulations, 10 technical data – Scilogex MS-S User Manual
Page 9

7
MS-(H)-S Analog (Hotplate) Magnetic Stirrer
cleaning.
• Unplug the power line when cleaning.
• Only use cleanser that we advised as below:
Dyes
Isopropyl alcohol
Construction materials
Water containing tenside /
isopropyl alcohol
Cosmetics
Water containing tenside /
isopropyl alcohol
Foodstuffs
Water containing tenside
Fuels
Water containing tenside
Table 3
• Before using other method for cleaning or decontamination,
the user must ascertain with the manufacturer that this
method does not harm or destroy the instrument.
• Wear the proper protective gloves during cleaning of the
instrument.
• The instrument must be cleaned and put it into the initial
packaging carton before sending to service for repair,
avoiding the contamination of hazardous.
• Use the instrument in a dry clean room and temperature
stable environment.
9 Associated standards and regulations
Construction in accordance with the following safety
standards:
EN 61010-1
UL 3101-1
CAN/CSA C22.2(1010-1)
EN 61010-2-10
Construction in accordance with the following EMC
standards:
EN 61326-1
Associated EU guidelines:
EMC-guidelines: 89/336/EWG
Machine guidelines: 73/023/EWG
10 Technical data
Items
Parameters
Voltage [VAC]
200-240 / 100-120
Frequency [Hz]
50/60
Power [W]
530(MS-H-S) / 130(MS-S)
Stirring point position
quantity
1
Max. stirring quantity (H2O) [L] 20