Caution – Scilogex D2012 Micro-Centrifuge User Manual
Page 17
![background image](/manuals/627983/17/background.png)
D2012 High Speed Mini Micro-Centrifuge
www.scilogex.com
Page 17
11.1.3 Notice
The centrifuge rotor can separate samples which density lower than 2.0 g/ml. If the separated samples density is over
2.0 g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm) = Maximum speed × (2.0 (g/ml) /Sample density (g/ml) )1/2
11.1.4 Autoclaving
The rotor is manufactured in high-strength aluminum alloy material and can be autoclaved: 121ć(1.0kg/cm2)20
minutes.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
Table 11.2 Cleaning and sterilizing conditions for tubes
O:Applicable X:Inapplicable
PA PC
PP
Cleaning
Cleaning fluids
Acidic(pH5 or lower)
X
X
X
Acidic(higher than pH5 )
O
O
O
Alkaline(higher than pH9 )
O
X
O
Alkaline(pH9 or lower)
O
O
O
Neutral(pH7)
O
O
O
Warm water(up to 70ć)
O
O
O
Ultrasonic cleaning
Neutral detergent(pH7)
O
O
O
Sterilization
Autoclaving
115ć(0.7kg/cm2)30minutes
O
O
O
121ć(1.0kg/cm2)20 minutes
X
O
O
126ć(1.4kg/cm2)15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
Formaldehyde
O
O
O
Materials
Conditions
PA:Polyallomer; PC:Polycarbonate; PP:Polypropylene
● The lid of the rotor is made of plastic, can not be high-pressure sterilization, only ordinary sterilization can be used.
CAUTION
ʽ
11.2.2 Cleaning PC tubes
PC material is low in chemical resistance against alkaline solutions. Avoid using neutral detergents with pH
higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according to the the
manufacturer’s instruction. Use detergent with its pH between 7 and 9.