● Read the user manual thoroughly to properly load and use rotor.
● Do not exceed the allowable maximum speed of rotor atube and adapters etc.. Ensure the allowable
maximum speed of adapters is lower than the rotor’s maximum speed.
11. Instructions for the Rotor and Tubes
11.1 Rotor instructions
11.1.1 Rotor structure
Figure11-1 The rotor structure
11.1.2 Available rotors and adapters
The rotor can be used as follows:
Table 11.1 Rotors and adapters
Rotor type
ID
code
Tubes
Adapters
Maximum
speed
(rpm)
Maximum RCF
(×g)
Allowable imbalance(*)
Imbalance
Volume
imbalance
A12-2
01
1.5/2.0ml
tube
15000
15100
2.0g/ tube
5mm/ tube
0.2ml
PCR tube
A02P2
15000
15100
0.5ml micro
tube
A05P2
15000
15100
CAUTION
ʽ
Locking nut
Fastener
Rotor lid
Rotor block
Shaft
Tube holes