Technical considerations – Bio-Rad Matrix-Prepared Surfaces User Manual
Page 4

3
© 2007 Bio-Rad Laboratories, Inc.
Technical Considerations
n
If your sample has salt or urea contamination, proceed directly
to Protocol 2. Otherwise, perform Protocol 1 first. If there is
insufficient signal after reading the arrays, a cleanup step is
recommended: Add 5 µl of 0.1% trifluoroacetic acid (TFA) to
the spot; do not agitate; remove the 0.1% TFA droplet after
30 sec; and add 5 µl 25% acetonitrile (ACN) and 0.1% TFA to
the spot. Allow to dry, and read the arrays again
n
With ProteinChip SEND ID arrays, it is essential that you mix
your sample with a solution of 50% ACN
and
0.2% TFA at a
1:1 (v/v) ratio before adding to the spot
n
Ideally, the final concentration of ACN after dilution is 25%;
a final concentration of greater than 40% ACN is not compatible
with the ProteinChip SEND ID array
Factors That Can Result in Weak Signal
n
We do not recommend washing on-spot (by pipetting the
sample up and down) because this can reduce signal
n
Weak signal can be improved by adding ACN to your sample
as described in Protocol 1
n
A predominant peak at mass-to-charge ratio (m/z) 211
indicates sodium contamination, and you should perform
sample cleanup as described in Protocol 2, step 5. This
contamination can suppress sample peaks and result in
weak signal