Bio-Rad trUView Cuvettes User Manual
Page 4

Technical Specifications
Outer Dimensions
12.5 x 12.5 x 45 mm
Light Path Height
8.5 mm
Pathlength
10 mm
Functional Wavelength Range
220–1,100 nm
Baseline Absorbance
< 0.3 AU @ 260 nm
Variability
< 15 mAU @ 260 nm
Functional Volume
> 50 µl
Chemical Compatibility
trUView cuvettes are resistant to all
chemical buffers typically used in most life
science laboratories. The trUView cuvettes
are not compatible with some solvents
such as chloroform and toluene.
Contaminants
All trUView cuvettes are guaranteed
RNase and DNase free.
5
Troubleshooting
Inaccurate Assay Values
– Ensure blank and sample are assayed using
the same cuvette orientation.
– For best results measure the blank and sample
in the same trUView cuvette.
– Inspect the cuvette for scratches or dirt on
optical windows.
– Ensure the light path height (z-dimension) of the
spectrophotometer is 8.5 mm. If it is 15 mm,
use the optional height adaptor.
– If cuvette has been reused, discard and use a
new cuvette. Multiple assays will degrade light
transmission properties of the cuvette.
– Visually check the sample has no air bubbles.
Air bubbles can be removed by gently tapping
the cuvette.
4
4006234A.qxp 1/15/2004 2:20 PM Page 4
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)