Fig. 7. Placing the focusing assembly

Fig. 8. Fitting the cathode assembly into

onto the anode of the chassis.

the cathode of the chassis.

3. Using a 10 ml syringe, add electrolyte solutions to the electrode

assemblies (Figure 9).

(Note: The electrode membrane storage solution from the starter kit can be
used. Alternatively fresh electrolyte solutions can be prepared)

I.

Add 6 ml 0.1 M H

3

PO

4

through the vent hole of the anode assembly

(red).

II.

Add 6 ml 0.1 M NaOH through the vent hole of the cathode

assembly (black).

Fig. 9. Adding electrolyte to the anode assembly.

4. Close the cooling block cover and tighten the screw.

5. Place the lid on the chassis.

6. Attach the power cord to the back of the MicroRotofor chassis and

connect it to an electrical outlet.

7. Connect the MicroRotofor cell to a vacuum source. We recommend

installation of a vacuum trap between the cell and the vacuum system.