Bio-Rad MicroRotofor™ Cell User Manual
Page 15

2. Remove the two screws that secure the Harvesting station to the vacuum
block and remove the positioning block, needle array, and needle array
holder (Figure 15).
Fig. 15. Disassembling the harvesting station (top) and harvesting station components
(bottom).
3. Clean and dry the positioning block.
4. Using a wash bottle, wash and rinse the individual needles in the needle
array.
5. Clean and dry the vacuum chamber.
6. Inspect and reposition the vacuum gaskets.
7. Reassemble the harvesting station.
13
10004399A.qxp 9/12/2005 2:08 PM Page 13