Very high-concentration extracts, with protein levels

above the upper limit of the standard curve, require that

lesser volumes of them be assayed. In such cases,

increase the amount of 0.12 N HCl (nominal) used so that

the total volume is 100 µl.

Add 3.5 ml of diluted dye reagent to each tube. Vortex gently.

Measure the absorbances at 595 nm (A

595

) after 5 minutes.

Use the standard curve generated in Step 4 to estimate the

concentration of the sample proteins from the A

595

measurements.

Representative Standard Curves.

Standard curves

were generated as described above with serial dilutions of
bovine gamma globulin in the three extraction solutions

14

Sequential Extraction Standard Curve

Bovine gamma globulin, in micrograms

0

50

100

150

200

250

300

12

Solution 3

Reagent 1

Solution 2

1

0%

0.6

0.4

0.2

0