Bio-Rad Aurum™ Vacuum Manifold User Manual
Page 7

Section 9
Troubleshooting Guide
Problem
Possible Cause
Possible Solution
Insufficient negative Open vacuum regulator
Close vacuum regulator
pressure in vacuum
by turning completely
manifold
clockwise
Manifold top not seated
Ensure uniform contact
properly on base
between manifold top
and base
Purification plate or CAP
Press down gently on
not fully seated
the corners of the plate
or CAP to seal
Unused CAP luer fittings
Close unused luer
not closed completely
fittings completely with
caps
Gasket residue preventing Rinse manifold with
even contact between
distilled water
manifold parts
Lower gasket misaligned
Press gasket back into
groove
Physical damage to
Replace vacuum
manifold or CAP (e.g.,
manifold or CAP
chips and cracks)
Purge step in purification
If closure of wells or
protocol
columns (e.g., with
sealing tape) restores
negative pressure, no
corrective action
needed
Excessive noise
Purification plate or CAP
Press down gently on
during vacuuming
not fully seated
plate or CAP to seat
Gasket residue preventing Rinse manifold with
even contact between
distilled water
manifold parts
Worn manifold gasket
Replace gasket
Problem
Possible Cause
Possible Solution
Incorrect negative
Vacuum source
Connect vacuum
pressure from
connected to left
to right side of vacuum
vacuum regulator
side of vacuum
Clogging of
Insufficient negative
See “Insufficient
purification plate
pressure in vacuum
negative pressure
or CAP columns
manifold
in vacuum manifold”
problem above
9