Bio-Rad BioLogic DuoFlow™ pH Monitor User Manual
Page 4

Life Science
Group
Web site www.bio-rad.com USA (800) 4BIORAD Australia 02 9914 2800 Austria (01)-877 89 01 Belgium 09-385 55 11 Brazil 55 21 507 6191
Canada
China 86-10-8201-1366/68 Denmark
Finland
France 01 47 95 69 65
089 318 84-177
Hong Kong 852-2789-3300 India (91-124) 6398112/113/114 Israel 03 951 4124 Italy 34 91 590 5200 Japan 03-5811-6270 Korea 82-2-3473-4460
Latin America
305-894-5950 Mexico 52 5 534 2552 to 54 The Netherlands
New Zealand 64-9-4152280
47-23-38-41-30
Russia 7 095 979 98 00 Singapore 65-2729877 Spain 34-91-590-5200 Sweden 46 (0)8-55 51 27 00 Switzerland 061-717-9555 United Kingdom 0800-181134
00-000 0000 Sig 0101
Bulletin 0000 US/EG Rev A
Bio-Rad
Laboratories, Inc.
Electrode Cleaning
A dirty pH electrode can lead to erroneous pH readings and shorten the lifetime of the electrode. Most contami-
nant’s will be either proteins that have adsorbed onto the glass membrane or buffer and salt crystals that have
blocked the pH probe pores.
Removal of protein deposits: Immerse the electrode in a 1% solution of proteases, such as trypsin or pepsin, in
0.1M HCl for no more than 5 minutes. Rinse with deionized water to remove all traces of HCl.
Removal of salt deposits: Immerse the electrode in 0.1M HCl for 5 minutes followed by 0.1M NaOH for 5 minutes
and then 0.1M HCl for 5 minutes. Rinse the electrode with deionized water to remove all traces of HCl.
Storing Your pH Electrode Between Uses
When pH readings are made infrequently (i.e., several days or weeks apart), the electrode should be stored in its
plastic bulb filled with electrode storage solution (available from VWR or Orion) or pH 4 buffer in 3.5 M potassium
chloride.
Do Not Allow the Electrode to Dry Out.
Trouble Shooting
Symptom
Possible Cause
Possible Solution
Noisy or unstable Readings
Air bubble on membrane
Remove air bubble
Drift
Solution temperature is changing
Allow all solutions to come to thermal equilibrium.
Leakage around pH probe
Loose nut
Tighten pH electrode nut
Leakage around pH probe
Missing or damaged o-rings
Replace o-rings
Leakage around fittings
Loose fitting
Tighten fitting
Leakage around fittings
Cracked or loose ferrule
Replace fitting
Technical Assistance
For additional help, contact your local Bio-Rad representative. In the United States, call Chromatography
Technical Support at 1-800-4BIORAD.
Ordering Information
760-2040
DuoFlow pH Monitor
760-2042
pH Electrode
760-2044
Flow Cell
750-0502
Signal Import Module
910-1828
o-ring, pH flow cell, size 2-009
910-3014
o-ring, pH electrode, size 2-018
760-2046
BioLogic pH Monitor tubing kit – for connection between Conductivity flow cell and pH flow cell inlet
includes
1 x 28 cm 1/16” OD x 0.02” ID orange PEEK, rated to 5000 psi
1 x 28 cm 1/16” OD x 0.03” ID green PEEK, rated to 3000 psi
4106525 Rev B
- BioLogic DuoFlow Pathfinder 80 System pH Monitor BioLogic DuoFlow Pathfinder 20 System pH Monitor BioLogic DuoFlow Maximizer 80 System pH Monitor BioLogic DuoFlow Maximizer 20 System pH Monitor BioLogic DuoFlow QuadTec 40 System pH Monitor BioLogic DuoFlow QuadTec 10 System pH Monitor BioLogic DuoFlow 40 System pH Monitor BioLogic DuoFlow 10 System pH Monitor