Description of parts, Parts layout – Amico Automatic Cryogenic Liquid Solenoid Manifold User Manual
Page 11

www.amico.com
11
OPERATING ALARM SYSTEMS
The manifold control cabinet contains the required circuitry to send a dry contact signal to the alarm unit when
a bank is empty and change-over occurs. The normally closed internal circuitry is designed to alarm when there
is an open circuit. The depletion of a bank triggers a relay, which renders the alarm circuit open and initiates the
alarm signal.
Description of Parts
Alarm Connection
Bank Transducer
Left Bank Gauge
Solenoid
Check Valves
Supply
Line Gauge
Power Supply
Bleed Valve 2X
Right Line Regulator
Bank Transducer
Solenoid
Operating Relief Valve
Relief Valve Vent
Right Bank Gauge
Left Line Regulator
Ball Valve 4X
Economizer Circuit
Provision for reserve
PARTS LAYOUT
Line Relief Valve
See also other documents in the category Amico Equipment:
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)