6.4 Medical Air Systems Glossary
9
Definitions
ICFM
•
Conditions at the inlet of the compressor prior to any restrictions or
temperature changes. (i.e., inlet filters, inter-coolers, discharge manifolds
etc).
Displacement
•
Theoretical physical volume of the air in the compressor chamber based
upon 100% pumping efficiency, with no allowances made for heat,
friction, clearances, or other losses in the compression cycle.
Peak Calculated
•
The maximum estimated demand a medical facility will require of a
Load (PCL)
medical air system.
•
Calculated at SCFM at 100 psig.
Oil-Free Compressor •
Compressor with oil within the compressor but absent from the
compression chamber by means of seals and visible vented distance
piece, and a dry compression chamber.
Simultaneous Demand •
Operating reference to a condition where all compressors (Lead and Lag
compressors) run simultaneously to satisfy demand in excess of lead
compressor(s)
capability.
Dew Point
•
Temperature and pressure at which medical air will condense water
vapor into liquid water within the medical air pipeline system.
Continuous Duty •
Air compressors which can operate continuously (24 hours per day) if
Rated
necessary but normally only operate on demand.
LPM
•
Abbreviation for Liters per minute; a measure of the flow rate of a gas.
Quadruplex System
•
Medical air system with four compressors (4), each sized for 33% PCL.
ACFM
•
Actual Cubic Feet per Minute is an expression of actual air volume,
generally corrected for and reference to a particular pressure.
Continuous Duty
•
Operational reference to compressors operating 24 hours a day,
continuously.
Duplex System
•
Systems comprised of two (2) compressors, each rated for 100% Peak
Calculated
Load
(PCL).
Desiccant Dryer
•
Consists of two towers, whereby the desiccant bed in one tower dries the
air stream while a purge of dry air regenerates the desiccant bed in the
other
tower.
NTP
•
“Normal Temperature and Pressure”, generally accepted as 70˚F (20˚C),
14.969 psi/29.92” Hg barometric pressure, and 36% relative humidity.
Dew Point/CO Monitor •
Monitors dew point and carbon monoxide levels in medical air.
Oil-Less Compressor •
Compressor with no oil contained within the compressor.
SCFM
•
Standard Cubic Feet Per Minute is an expression of air at NTP.
Triplex Systems
•
Medical air system with three compressors (3), each sized for 50% PCL.