Monarch Instrument ACT-1B Series User Manual
Page 3
OVERVIEW ..................................................................................... 1
INSTALLATION and POWER .......................................................... 1
Installation ............................................................................... 2
Power ...................................................................................... 2
USB Programming Cable and Software ................................ 3
SENSOR CONNECTIONS ............................................................. 3
OUTPUT OPTIONS ......................................................................... 4
Current Output Option (IO) ...................................................... 5
Analog Output Option (AO) ...................................................... 6
Pulse Repeater Option (PO) ................................................... 6
SPECIFICATIONS ........................................................................... 7
OPTIONS ........................................................................................ 8
ACCESSORIES / SENSORS .......................................................... 9
APPENDIX A - Serial Programming Commands ......................... 10
TABLE OF CONTENTS
TABLE OF CONTENTS
@CH_A/LOEND = 12 (or 1_SEC, HALF) Sets low end time. This allows a
min reading of 5 RPM, 60 RPM, or 120
RPM.
@CH_A/GATE
Shows Gate Speed. (Default is 12)
@CH_A/GATE = STD (1/100 Second) or FAST (1/1000 second). Sets
Gate Speed (Default is 1/100)
@DECPT
Shows the number of decimal places
displayed
@DECPT = NONE, 1, 2, or 3
Sets the maximum number of decimal
places
@DAC1/ FSCAL
Shows Analog Out Full Scale
@DAC1/FSCAL = xxx.xx
Sets the Reading value that the Analog
output will output Full Scale (5V or
20mA). Depends on TYPE.
@DAC1/0SCAL
Shows Analog Out Zero Scale
@DAC1/0SCAL = xxx.xx
Sets the Reading value that the Analog
output will output Zero Scale (0V or
4mA). Depends on TYPE. (Default is
0.00)
@OUTPT
Shows pulse output polarity
@OUTPT = POS or NEG
Sets pulse output polarity
@DISPR
Shows Display Update Rate
@DISPR = HALF or 1_SEC or 1.5_S. This sets the maximum display
update rate to one half a second, 1 second
or 1 ½ seconds between updates.
@SERNO
Shows unit Serial Number
11