Specifications – VICI D-4-I Instruction Manual User Manual
Page 10

8
Specifications
Mains (line):
115/230 V~50/60 Hz, 175 VA
Fuse:
2 A, time-delay, 5 x 20 mm
Pressure:
6.9 kPa (1 psi) operating, 6.9 MPa (1000 psi) max.
working
Maximum temperature 400°C
Heater power:
60 W max., 48 V, PWM
Sensitivity
10X range:
1.0 V/nA
1X range:
0.1 V/nA ± 1%*
Range
10X:
10 nA full-scale
1X:
100 nA full-scale*
Noise
1X:
20 fA/
√
Hz (referred to the input), 0.1 - 10 Hz*
10X:
5 fA/
√
Hz (referred to the input), 0.1 - 10 Hz
Risetime:
10 msec, 10% = 90%*
Output impedance:
< 1
Ω
, attenuated
1 K
Ω
, unattenuated, 1 V, 10 V
*Controller only, PDHID mode, 10 V unattenuated output