HF scientific CLX Online Residual Chlorine Monitor (Old) User Manual
Page 33

Return to Normal Operation
Press the SERVICE button to return sample flow to the system. Check for leaks. If a leak
occurs press SERVICE again, repair leak and try again. Once the system is operating
correctly, return or replace reagents and press PRIME and then one time to restart
reagent flow. The system will automatically return to normal operation.
Notes: Tubes may darken due to contact with the reagent. This condition does not affect the
performance of these parts.
TUBING "D"
TUBING "C"
INLET
CHECK
VALVES
TUBING "B"
REAGENT
BOTTLE
CAP
PUMP
HAMMER
THUMB
SCREW
REAGENT
PUMP
ASSY
INLET
OUTLET
FLOW
y
CHECK VALVE DETAIL
OPTICS SYSTEM
KNURLED TOP
CUVETTE
RETAINING
O-RING
CLX (9/07)
Page 27
REV 3.1