beautypg.com

0 troubleshooting, 1 clx fault detection – HF scientific CLX Online Residual Chlorine Monitor (Old) User Manual

Page 29

background image

CLX (9/07)

Page 23

REV 3.1

9.0 Troubleshooting

9.1 CLX Fault Detection
The CLX performs continuous diagnostic monitoring. In the CLX, there are 4 severity
levels of fault detection. Level 4, 3 & 2 will allow normal operation, but warn of the
problem. Level 1 is an instrument failure and the instrument will not operate. Any faults
are displayed in a queue form in the bottom row of the LCD.

A level 4 fault is simply a screen indication that one of the alarm levels has been
activated. This fault level will not affect the 4-20 mA and will only affect the alarm
activated. The sample back light blinks at a rate of once every 4 seconds.

A level 3 fault indicates a failure or a problem that usually can be corrected by the
operator. Refer to the chart below. If any of these errors occur, the instrument will still
display readings and probably will operate correctly. These faults will self-clear when the
problem is corrected. If any of these faults occur, they will affect the 4-20mA and any
alarm set for fault detection (Error). The sample back light blinks at a rate of once every
2.5 seconds.

Message

Description of Fault

Corrective Action

MA

4-20 mA enabled & loop open

Check 4-20 mA wiring or turn off 4-

20mA if not used

CAL

Calibration invalid –not accepted

Recalibrate if needed

WATR

No water flowing

Check water flow

FAST

Intake water flow too fast

Set flow rate (see section 4.2.1)

SLOW

Sample cuvette filling too slowly

Check water flow

PURG

Sample cuvette has slow purge

Check drain lines

NPRG

Sample cuvette not purging

Check drain lines

H2O

CLX cannot determine water level

Check water flow

ISOL

Problem with intake solenoid

Check wiring, check for clogged solenoid

PSOL

Problem with purge solenoid

Check wiring, check for clogged solenoid

RGNT

Problem with reagent

Check reagent and lines

GLAS

Dirty cuvette

Replace or clean cuvette

WCAL

Water Level Calibration Invalid

Clear fault (see section 9.2)

REPL

Reagents Expired

Replace reagents and press PRIME

A level 2 fault indicates a severe problem that will usually require technical assistance
from HF scientific customer service (see section 9.4). The queued display will show
POST. If this fault occurs it will affect the 4-20mA and any alarm set for fault detection
(Error). The sample back light blinks at rate of once every 1 second.

A level 1 fault is a system fault. This is NOT a problem that the operator can correct, and
the unit must be returned to the factory for service (see section 9.4). These failures consist
of failures in the CPU, A/D, EEPROM or other devices internal to the instrument. The
queued display will show FAIL, the upper display is a five digit code. If this fault occurs,
it will affect the 4-20mA and any alarm set for fault detection (Error). The instrument will
not operate with this fault. The sample back light blinks at rate of once every 0.4 seconds.

If any fault conditions occur, the message indicating the fault will be shown on the lower

row of the display.