3 Installation

892 Professional Rancimat

■■■■■■■■

17

Figure 8

Mounting the air tube: correct - incorrect

WARNING

The foam barrier can melt if it projects too deeply into the heating
block.

Ensure that the foam barrier is at least 7 cm above the base of the
reaction vessel.

Figure 9

Mounting the foam barrier

3.2.4

Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them
in the 892 Professional Rancimat and establish the tubing connections
(see Chapter 2.1, page 6) as follows:

1 Inserting the measuring vessel

Fill distilled water into the measuring vessel.

Place the measuring vessel cover onto the measuring vessel.

Insert the measuring vessel into the openings provided on the 892
Professional Rancimat. While doing so, carefully guide the con-
nector plug into the electrode connector.

Connect the white silicone tubing to the M8/olive tubing adapter
of the measuring vessel cover.