Metrohm MagIC Net Empower Driver User Manual

Page 11

background image

▪▪▪▪▪▪▪

9

8.

Click on Add….

Note: If several systems are added, both the IP address and the MAC address have to differ.

9.

Enter the following parameters:

IP address:

000.000.000.xxx, (where x = 001-255)

e.g. start with 000.000.000.001

MAC address:

00-00-00-00-00-xx, (where xx = 00-FF)

e.g. start with 00-00-00-00-00-01

Instrument type:

Select "Metrohm IC"

This includes all implemented Metrohm instruments, also Sample Processors and
Dosinos.

Serial Number / Unique Name: These are the actual settings used to detect the
instrument on the USB bus and must meet the following requirements:

The serial number must be the last five characters (including leading 0s) printed on the
serial number sticker found on the back of the instrument.

The IC serial number must be prefixed with "IC="; the Sample Processor must be
prefixed with "SP=" (excluding double quotes).

The IC and SP serial numbers must be separated by a space.