3.24 Conditioning
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
940 Professional IC Vario ONE/SeS/PP/Prep 2 (2.940.1520)
3.24
Conditioning
After installation, after each time the instrument is switched on and after
each time the eluent is changed, the system has to be conditioned with
eluent long enough that a stable baseline is attained.
NOTE
The conditioning time can lengthen considerably after changing the
eluent.
Conditioning the system
1 Preparing the software
CAUTION
Ensure that the configured flow is not higher than the flow permit-
ted for the corresponding column (refer to the column leaflet and
the chip data record).
■
Start the MagIC Net computer program.
■
Open the Equilibration tab in MagIC Net: Work-
place
▶ Run ▶ Equilibration.
■
Select (or create) a suitable method.
Also see: MagIC Net tutorial and online help.
2 Preparing the instrument
■
Ensure that the column is inserted correctly in relation to the flow
direction marked on the sticker (arrow has to point in the direc-
tion of flow).
■
Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.
3 Starting equilibration
■
Start the equilibration in MagIC Net: Workplace
▶ Run ▶ Equi-
libration
▶ Start HW.