Metrohm 930 Compact IC Flex Oven/SeS/Deg User Manual
Page 99

■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
■■■■■■■■
91
Problem
Cause
Remedy
Suppressor blocked.
■
Regenerate the suppressor (see Chapter
5.12.3.2, page 79).
Note: 6.2821.180 pump tubing connection
with filter must be used.
Guard column – Blocked.
Replace the guard column (see Chapter 3.19,
page 46).
Separation column –
Blocked.
■
Regenerate the separation column (see
Chapter 5.16.4, page 88).
■
Replace the separation column (see "Con-
necting the separation column", page 50).
Note: Samples should always be microfiltered .
Injection valve – Valve
blocked.
Have the valve cleaned (by a Metrohm service
technician).
The retention times
in the chromato-
gram have changed
unexpectedly.
Separation column –
Diminished separating effi-
ciency.
■
Regenerate the separation column (see
Chapter 5.16.4, page 88).
■
Replace the separation column (see "Con-
necting the separation column", page 50).
The eluent contains gas
bubbles.
■
Ensure that the connectors for the eluent
degasser are firmly connected .
■
Deaerate the high-pressure pump (see Fig-
ure , page 45).
High-pressure pump –
Defective.
Request Metrohm Service.
Peak areas are
lower than
expected.
Sample – There is a leak in
the sample path.
Find and fix the leak in the sample path.
Sample – The sample path
is blocked.
Find and fix the blockage in the sample path.
Sample – The sample loop
is not (completely) filled.
Adjust the transfer time for the sample (see
"Determining the transfer time", page 86).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
MCS – The CO
2
suppressor
is not connected.
Connect the CO
2
suppressor.
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean the contact surfaces of the column chip
with alcohol.