14 installing the detector – Metrohm 940 Professional IC Vario ONE/ChS User Manual
Page 46
3.14 Installing the detector
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
940 Professional IC Vario ONE/ChS (2.940.1200)
4 Connect the FEP tubing (6.1805.120) from the bottle with the rinsing
solution with the adapter Dosino Port 4/M6 to the Port 4 of the Dos-
ino.
If you would like to use a separate Dosino for the rinsing solution, you will
need the following additional accessories:
■
Dosino (2.800.0010)
■
Dosing unit 50 mL (6.3032.250)
■
Accessory kit: Flex/Vario: ChS (6.5000.030)
■
IC equipment: Dosino Regeneration (6.5330.190)
Connect the rinsing solution capillary with the Dosino and the bottle with
the rinsing solution as described in chapter Connecting the regeneration
solution to the Dosino, page 35.
3.14
Installing the detector
The 940 Professional IC Vario provides enough space for two detectors
and additional accessories in the detector chamber. The detectors are
available as accessories and are supplied with separate manuals.
Placing the detector in the instrument
Follow the instructions in the chapter Inserting the detector in the manual
for the detector.
Connecting the detector to the eluent path
Accessories
For this step you need the following accessories:
■
Coupling (6.2744.040)
■
2 × pressure screw (6.2744.010)
1 Connect the out capillary of the suppressor and the detector input
capillary to one another using a coupling (6.2744.040) and two short
pressure screws (6.2744.070).