Metrohm 807 Dosing Unit User Manual
Page 50
6.2 Dosing accuracy
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
807 Dosing Unit
Table 4
Permissible limit values as per ISO/EN/DIN 8655-3
Cylinder
volume
(mL)
max. systematic
Measurement deviation
max. permissible
Measurement deviation
2
± 0.5 %
± 10 µL
± 0.1 %
± 2 µL
5
± 0.3 %
± 15 µL
± 0.1 %
± 5 µL
10
± 0.2 %
± 20 µL
± 0.07 %
± 7 µL
20
± 0.2 %
± 40 µL
± 0.07 %
± 14 µL
50
± 0.2 %
± 100 µL
± 0.05 %
± 25 µL
The Metrohm agents worldwide offer the possibility of on-site dosing unit
inspections and certifications with respect to accuracy. We recommend
that an accuracy inspection be performed when the dosing cylinders and
dosing pistons of a dosing unit are replaced.