Page 31: Dosing cylinder damaged

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

807 Dosing Unit

■■■■■■■■

23

Figure 16

Cylinder unit on the distributor

Removing the centering tube

Remove the centering tube as follows:

1 Remove the cylinder unit, comprised of centering tube, black cylinder

base, valve disc, dosing cylinder and dosing piston, from the distribu-
tor.

2 Carefully pull out the centering tube on the black cylinder base in an

upward direction.

You will now see the dosing cylinder on the black cylinder base with
the integrated valve disk.

You can now inspect the condition of the cylinder and the piston.
The glass cylinder should not exhibit any signs of corrosion. The plas-
tic coating PTFE of the piston should not be damaged in any way.

Figure 17

Dosing cylinder damaged