6 Appendix
807 Dosing Unit
■■■■■■■■
41
2
1
Figure 30
Data chip and contact pin
1
Contact surfaces
on the 800 Dosino
2
Data chip with contact pins
on the 807 Dosing Unit
6.2
Dosing accuracy
Every dosing unit is subjected to a strict quality inspection prior to ship-
ment. Every dosing unit is issued a quality certificate attesting conform-
ance with the quality criteria of Metrohm.
6.2.1
Typical measurement deviation
The accuracy of dosing units can be seen in the following table. The val-
ues listed are to be regarded as typical values which can be achieved with
a 700 Dosino or an 800 Dosino.
Table 3
Typical measurement deviation of Metrohm dosing units
Cylinder volume
max. systematic deviation
2 mL
± 6 µL
5 mL
± 15 µL
10 mL
± 20 µL
20 mL
± 30 µL
50 mL
± 50 µL
6.2.2
The ISO/EN/DIN standard 8655-3
The Metrohm dosing units fulfil the requirements of the ISO/EN/DIN
standard 8655-3 Volume measurement instruments with pistons –
Part 3: Piston burets. Metrohm guarantees that its dosing units are in
compliance with the following limit values at the time of shipment: