Regenerating the suppressor in case of high counterpressure
If excessive counterpressure is observed in one or more of the suppressor units,

proceed as follows:

1 Disconnect suppressor from IC system

• Disconnect suppressor from separating column and detector.

2 Regenerate suppressor

• Connect the inlet capillary 25 marked "H

2

SO

4

" to the 709 IC Pump using a

6.2620.060 or 6.2744.040 coupling.

• Adjust the 709 IC Pump flow rate to 0.5 mL/min and rinse the suppressor

unit with 1 mol/L H

2

SO

4

for 5 to 10 min.

• As the pressure drops, slowly increase the 709 IC Pump flow rate to

2 mL/min. The maximum pressure must not exceed 2 MPa (20 bar).

• Switch suppressor to the next position using the key.
• Connect the inlet capillary 28 marked "H

2

O" to the 709 IC Pump using a

6.2620.060 or 6.2744.040 coupling.

• Adjust the 709 IC Pump flow rate to 0.5 mL/min and rinse the suppressor

unit with distilled H

2

O for 5 to 10 min.

• As the pressure drops, slowly increase the 709 IC Pump flow rate to

2 mL/min. The maximum pressure must not exceed 2 MPa (20 bar).

3 Connect suppressor to IC system

• Reconnect the suppressor to the IC system and the 753 Suppressor

Module. If the pressure problems persist, replace the suppressor rotor

(see section 4.2.7).

Supplement to «Cleaning the suppressor»

p. 30

Supplementary to the cases listed in section 4.2.6 which may necessitate cleaning of

the suppressor or replacement of connection piece 36, please also note the following

case:

• High counterpressure on the suppressor connection

tubings

Changes in standard equipment

p. 37

2 × 6.2744.010 PEEK Compression fitting (instead of 1 × 6.2744.010)

Set of 5

2 × 6.2128.180 Connecting cable (instead of 1 × 6.2128.090)

Connecting cable 762 – 752/753/754

2 × 6.2821.100 Filter unit PEEK 2 µm (new)

To avoid contamination due to abrasive particles of piston

seals

New Options

p. 38

6.2143.220 Connecting

cable

Connecting cable 732 IC Detector – 753 Suppressor Module –

754 Dialysis Unit

753 Suppressor Module, 8.753.1023