3. Troubleshooting
6.5613.000 Titration Equipment
24
3. Troubleshooting
Problem
Corrective action
Sticking AgCl precipitates.
•
Add anti-flocculating agent (e.g.
Polyviol®).
Insufficient cleaning.
•
Feed the 703 Ti Stand via separate
power adapter (see page 26).
•
Place the inserts (electrode, buret
tip) directly before the rinsing noz-
zles.
Rinsing solution is dripping.
•
Work with only 1 rinsing nozzle.
Place the inserts (electrode, buret
tip) as closely as possible to this
rinsing nozzle.