1. Assembly

6.5613.000 Titration Equipment

20

1. Assembly

Use the ECO titration equipment together with the 703 Ti Stand. It

is best to feed the 703 Ti Stand with a separate 11 V power

adapter (and not directly from the Titrator).

Ordering numbers for the power adapter:

220...240 V

6.2117.030

110...117 V

6.2119.030

1.1 Assembly for KF titrations

see picture on the front page.

1.

Fix titration vessel on the stand rod.

2.

Set the 6.2740.030 rinsing nozzle into the left rear aperture:

Put 6.2730.030 nipple, then E.301.0022 O-ring onto the

rinsing nozzle and screw it into the aperture.

3.

Cut 2 cm of 6.1543.120 aspirating tip and place it in the right

front aperture: Set 6.2730.030 nipple, then E.301.0022 O-

ring onto the aspirating tip and screw it into the aperture.

4.

Close the small apertures with 6.2730.030 stoppers.

Close the big aperture to the right with 6.2730.060 nipple

and 6.2709.090 stopper.

5.

Screw 6.2730.070 stopper into the left aperture and insert

6.0338.100 KF electrode without O-ring.

6.

Fill drying tube with drying agent (e.g. molecular sieve, or-

dering number 6.2811.000) and screw it into the middle ap-

erture using 6.2730.070 stopper.

7.

Set 6.2730.020 septum stopper into the middle aperture in

the front.

8.

Remove stopper of the rear middle aperture and insert buret

tip with titrating agent.

9.

Place stirring bar in the titration vessel and fix the vessel to

the lid. Take care that the 6.1454.030 gasket is straight.

Adjust the height of the titration vessel.

10. Set 6.1808.100 distributor onto the titration vessel lid and fix

it at the stand rod.

11. Connect the rinsing nozzle with the left aperture of the dis-

tributor using 6.1805.010 tubing.

Close the 2 remaining apertures of the distributor with

6.1446.040 stoppers.

12. Connect the tubing of the rinsing solution at the side of the

distributor.