background image

4.17 Initial start-up

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

940 Professional IC Vario ONE/Prep 1 (2.940.1110)

4 Deaerating the high-pressure pump

Use the syringe to aspirate eluent until there are no more air

bubbles in the eluent.

5 Setting the contact pressure of the peristaltic pump

Set the contact pressure of the peristaltic pump (see "Setting the
contact pressure correctly", page 22)
.

6 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.