Specifications, Connection – Yamato Scientific External output controller type Ⅳ CR User Manual
Page 2
1. Specifications
Time-up Output
•
Output at the automatic stop and at the end of program
•
C contact (Relay contact)
•
Contact capacity
:AC250V 3A
:DC30V 3A
•
Joint
:M4 Screw terminal (Common)
Alarm Output
•
Output at the occurrence of error
•
C contact (Relay contact)
•
Contact capacity
:AC250V 3A
:DC30V 3A
•
Joint
:M4 Screw terminal (Common)
Temperature Output
(Analog)
•
Voltage output(DC) according to the measured temperature
•
See table 1 in the next page about output
(0
℃ at 0V)
•
Resolution
:0.1
℃
•
Load impedance
:100K
Ω or more
•
Joint
:M4 Screw terminal (Common)
2. Connection
1. Connect firmly the terminal to use (Outputs which was not specified at the order are
unusable)
2. To use the time-up output or the alarm output;
Connect to COM (C) and NO in case the contact is “Closed” at output.
Connect to COM (C) and NC in case the contact is “Opened” at output.
3. when you use the temperature output, employ a shielded line as a connecting cable for
noise prevention, and cut it as short as possible.
Lengthwise Arrangement
Crosswise Arrangement