beautypg.com

Calibration, combustible and toxic versions – RKI Instruments Eclipse User Manual

Page 14

background image

Eclipse Detector Head Operator’s Manual

14

Calibration, Combustible and Toxic
Versions

This section describes how to calibrate the combustible gas and toxic gas versions of the
Eclipse. It includes procedures to prepare for calibration, enter the Calibrate menu, set the
zero reading, set the response reading, and return to normal operation. See the next
section, Calibration, Oxygen Version, for instructions to calibrate an oxygen Eclipse.

WARNING: The Eclipse is not an active gas monitoring device during the calibration

procedure. The 4-20 mA output signal will “freeze” at the level it was at
just before entering the Calibration Menu. It will not indicate current
readings until the Eclipse is in normal operation again.

Preparing for Calibration

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, zero air cylinder, fixed flow regulator, calibration cup for
the appropriate sensor, and a short piece of sample tubing to connect the
regulator to the calibration cup.

NOTE:

The following procedure assumes that hazardous gases are present. If a fresh air
environment can be verified, applying zero air to the sensor is not necessary
when setting the zero reading.

1.

Connect the regulator to the calibration cup using the sample tubing.

2.

Install the calibration cup over the detector, then screw the regulator into the zero air
calibration cylinder.

3.

Allow zero air to flow over the sensor for 1 minute then proceed.

Entering the Calibration Menu

1.

Press and release (cover and uncover) the optical switch with your finger 3 times and
then press and hold. After each press and release, a number will display in the upper
right of the LCD indicating how many times the sensor has been pressed and
released. If you press and hold your finger over the sensor, an “H” will appear in the
upper right indicating that you are pressing and holding.

2.

After 5 seconds, the following will display: CAL MODE HOLD SW. If you hold your
finger over the sensor for 5 more seconds, the following will display: RELEASE to
ZERO. Release your finger at this time and the current reading will display with the
message ZERO in the upper left of the display and an “R” in the upper right
indicating that the Eclipse is ready for a zero operation.

NOTE:

If there is no switch activity for 60 seconds while the screen indicates that the unit
is ready for a zero operation, the unit will return to normal operation.