RKI Instruments 65-2615RK-05-04 User Manual
Page 32

27
65-2615RK-05-04 M2 Transmitter Operator’s Manual
4.
When the desired target gas is on the display, press and release the ENTER button.
The display will ask SAVE IT? YES/NO.
5.
To discard the gas type change, press and release the DOWN/NO button. The display
will ask DO OVER? YES/NO. Press the DOWN/NO button. The display will ask
ABORT? YES/NO
. Press the UP/YES button. The M2 will return to normal operation
without changing the gas type.
To save the gas type, press and release the UP/YES button. The display will indicate
Gas Type Updated
, the display will indicate the target gas and the detection range for
a few seconds, and then the M2 will enter Configuration Mode automatically so that
the parameter settings in Configuration Mode can be verified or changed.
6.
While in Configuration Mode press and release the ENTER button to accept the
displayed parameter setting and move to the next one. Use the UP/YES and DOWN/
NO buttons to adjust a setting. See “Chapter 5: Configuration Mode” on page 23 for a
complete description of Configuration Mode.
7.
When you have reviewed and accepted the last item in Configuration Mode by
pressing the ENTER button, display will ask SAVE IT? YES/NO.
8.
To save the configuration settings, press and release the UP/YES button.The display
will indicate Config Saved and the M2 will begin its warm-up sequence.
To discard the settings and review them again, press the DOWN/NO button. The
display will ask DO OVER? YES/NO. Press the UP/YES button and the display will
indicate Re-do Config, then display the target gas and the detection range for a few
NO
0 - 100 ppm
HF
0 - 9.00 ppm
HCN
0 - 15.0 ppm
HCL
0 - 15.0 ppm
H2SE
0 - 5.00 ppm
GEH4
0 - 1.50 ppm
F2
0 - 3.00 ppm
CLO2
0 - 1.00 ppm
B2H6
0 - 5.00 ppm
ASH3
0 - 1.50 ppm
SO2
0 - 6.00 ppm
NH3
0 - 200 ppm
NH3
0 - 75.0 ppm
CL2
0 - 3.00 ppm
CL2
0 - 10.0 ppm
Table 6: Gas Types
Gas Type
Choices
Detection Range