Maintenance – RKI Instruments 65-2516RK User Manual

Page 12

background image

8 • 65-2516RK Oxygen Transmitter

12. Turn the regulator’s on/off knob clockwise to close it.

13. Unscrew the calibration cup from the detector.

14. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave

the sample tubing connected to the regulator and the calibration cup.

15. Store the components of the calibration kit in a safe and convenient place.

16. Remove the voltmeter leads from the test points.

17. Secure the junction box cover to the junction box.

Maintenance

This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum
performance of the oxygen transmitter. It includes daily, monthly, and quarterly
procedures.

Daily

Verify a display reading of 20.9% oxygen at the controller. Investigate significant changes
in the reading.

Monthly

This procedure describes a test to verify that the oxygen transmitter responds properly to
oxygen deficiency.

NOTE:

Performing a response test on the oxygen transmitter may cause alarms. Be sure
to put the controller into its calibration program or disable external alarms before
performing this test.

Preparing for the response test

1.

Verify that the display reading for the channel you are testing is 20.9% oxygen.

If the display reading is not 20.9% oxygen, set the fresh air reading of the transmitter
as described in “Start Up” on page 7 of this manual, then continue this procedure.

2.

Set a voltmeter to measure in the millivolt (mV) range.

WARNING: Do not remove the detector housing cap or junction box cover while the

circuits are energized unless the area is determined to be non-hazardous.
Keep the detector housing cap and junction box cover tightly closed during
operation.

3.

Remove the junction box cover, then plug the voltmeter leads into the test points on
the amplifier. Plug the positive lead into the test point labeled TP+; plug the negative
lead into the test point labeled TP-.

Performing the response test

1.

Exhale into the bottom of the oxygen detector.

2.

Stop exhaling into the bottom of the detector, then verify that the reading on the