65-2507RK Oxygen Transmitter • 13
4.
If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).
5.
Turn the regulator knob clockwise to close the regulator.
6.
Remove the regulator from the calibration cylinder.
7.
Unscrew the calibration cup from the oxygen detector.
NOTE:
For convenience, leave the components of the calibration kit connected by the
sample tubing.
Returning to Normal Operation
1.
Remove the voltmeter leads from the amplifier test points.
2.
Secure the junction box cover to the junction box.
3.
When the display reading rises above the decreasing alarm setpoint, return the
controller to normal operation.
NOTE:
If the oxygen reading is below the decreasing alarm setpoint when you return to
normal operation, the controller will initiate the appropriate alarm condition.
4.
Verify that the display reading increases and stabilizes at 20.9% volume.
5.
Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 6 lists replacement parts and accessories for the oxygen transmitter.
Table 6:
Parts List
Part Number
Description
06-1248RK
Sample tubing (3/16 in. x 5/16 in.; specify length when ordering)
18-0405RK-01
Junction box (does not include lid)
18-0406RK
Junction box lid
57-1061RK
Amplifier (specify target gas when ordering)
65-2506RK
Oxygen detector, capillary type, conduit-mounting
65-2507RK
Oxygen transmitter (includes detector and amplifier)
71-0080RK
65-2507RK Oxygen Transmitter Operator’s Manual (this document)
81-0076RK-01
Zero air calibration cylinder (34 liter)
81-0078RK
Calibration cylinder (100% nitrogen, 17-liter)
81-0078RK-01
Calibration cylinder (100% nitrogen, 34-liter)
81-1050RK
Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter steel
calibration cylinders
81-1111RK
Calibration cup