RKI Instruments 65-2507RK User Manual
Page 14

10 • 65-2507RK Oxygen Transmitter
7.
Connect the detector leads to the amplifier’s detector terminal strip as shown in
Table 3 below and Figure 3 on page 6 of this manual.
8.
Turn on power to the controller.
9.
Turn on the controller.
10. Calibrate the replacement detector as described in “Calibration” on page 12 of this
manual.
Replacing the Amplifier
1.
Turn off the controller.
2.
Turn off power to the controller.
3.
Remove the junction box cover.
4.
Disconnect the detector leads from the detector terminal strip.
5.
Disconnect the wiring that connects the oxygen transmitter to the controller from the
amplifier’s interconnect terminal strip.
6.
Unscrew and remove the two screws that secure the amplifier to the junction box.
The screws are at the top left and bottom right of the amplifier.
7.
Remove the amplifier.
8.
Place the new amplifier in the same position as the amplifier you removed in the
previous step.
9.
Use the two screws you removed in step 5 to secure the amplifier to the junction box.
10. Reconnect the wiring that connects the controller and oxygen transmitter to the
amplifier’s interconnect terminal strip as shown in Table 4 below and Figure 3 on
page 6 of this manual.
Table 3:
Reconnecting the Oxygen Detector
to the Amplifier
Detector Lead
Detector Terminal Strip
White
OXY W
Green
OXY G
Table 4:
Reconnecting the Oxygen Amplifier
to the Controller
Amplifier Interconnect
Terminal Strip
Controller
Transmitter Terminal
Strip
4/20
4 - 20 (FB)
24V
+ (11 VDC - 30VDC)