20 • 65-2485RK-02 Multi Point Detector

Returning to Normal Operation

1.

Unscrew the calibration cup from the oxygen detector.

NOTE:

For convenience, leave regulator and calibration cup connected by the sample
tubing.

2.

When the controller display reading rises above the alarm points, return the
controller to normal operation.

NOTE

: If you do not allow the gas reading to rise above the alarm points, then unwanted

alarms may occur.

3.

Verify that the controller display reading increases and stabilizes at 20.9% oxygen.

4.

Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 5 lists replacement parts and accessories for the multi-point detector.

Table 3: Parts List

Part Number

Description

06-1248RK-03

Calibration kit sample tubing, 3 ft. length

07-0033RK

Detector housing cap gasket (for H

2

S, O

2

and CO detectors)

07-0203RK

Rubber retaining boot (for CO and H

2

S detectors)

14-2101RK

Spacer between H

2

S sensor and rubber boot

33-7101RK

Charcoal filter disk (for CO detector)

65-1025RK

Oxygen replacement plug-in sensor

65-2423RK-05

H

2

S replacement detector assembly (includes sensor)

65-2433RK-05

CO replacement detector assembly (includes sensor)

65-2485RK-02

Multi-point detector assembly, O

2

/H

2

S/CO

65-2514RK

Oxygen replacement detector assembly (includes sensor)

71-0237RK

65-2485RK-02 Multi Point Detector Operator’s Manual (this document)

81-0064RK-01

Calibration cylinder, 50 ppm CO in air, 34 liter steel

81-0064RK-03

Calibration cylinder, 50 ppm CO in air, 103 liter steel

81-0076RK

Zero air calibration cylinder, 17 liter steel

81-0076RK-01

Zero air calibration cylinder, 34 liter steel

81-0076RK-03

Zero air calibration cylinder, 103 liter steel