Overview, Specifications – RKI Instruments 65-2336RK User Manual
Page 5

65-2336RK CO Transmitter • 1
Overview
This instruction manual describes the 65-2336RK carbon monoxide (CO) transmitter. This
manual also describes how to install, start up, configure, maintain, and calibrate the
transmitter when using it with a gas monitoring controller. A parts list at the end of this
manual lists replacement parts and accessories for the CO transmitter. See the controller
operator’s manual for information specific to the controller.
Specifications
WARNING: Do not use this product in a manner not specified in this instruction
manual.
Table 1 lists specifications for the CO transmitter.
NOTE:
The following symbol on the detector label is a caution to the user to refer to this
documentation for installation and operation instructions:
WARNING: When using the 65-2336RK, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 65-2336RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2336RK as described in this manual.
Table 1: Specifications
Target Gas
Carbon Monoxide
Area Classification
Explosion-proof for Class I, Groups B, C, and D
Temperature Code
T6
Installation Category
Installation Category 1. Signal level, special equipment or parts of
equipment, telecommunication, electronic, etc., with smaller
transient overvoltages than Installation Category (Overvoltage
Category) II (ref. IEC 664).
Input Voltage
11 VDC-30 VDC
Sampling Method
Diffusion
Signal Output
4 to 20 mA
Detection Range
0 to 300 PPM (parts per million)
Accuracy
± 5% of reading or ± 5 ppm CO (whichever is greater)
CO Detector Signal Output
0.3 mA at 0 ppm CO nominal
2.5 mA at 300 ppm CO nominal
Response Time
90% in 30 seconds
Operating Temperature
23°F to 104°F (-5°C to 40°C)
!